medical Class II Updated 2020-01-15

Baxter Healthcare Corporation recalls ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N

Recalled Product

ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N8375K, 7N8376K, 7N8377, 7N8378, 7N8390, 7N8391, 7N8399 - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use. This device is an in-line injection site which can be connected to standard male Luer adapters (e.g., syringes or sets) for the continuous or intermittent fluid administration or the withdrawal of fluids.


Hazard / Issue

Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: PRODUCT CODE UDI 7N8300 85412091839 7N8301 85412475387 7N8301K 85412499635 7N8310 85412476728 7N8330K 85412613499 7N8332K 85412599625 7N8334K 85412613482 7N8370K 85412613505 7N8371 85412091822 7N8375K 85412613475 7N8376K 85412599601 7N8377 85412091808 7N8378 85412091792 7N8390 85412475417 7N8391 85412475400 7N8399 85412091785 7N8399K 85412478227 All Lots
View official government recall