medical Class II Updated 2020-01-22

Trilliant Surgical, LLC recalls Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usa

Recalled Product

Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant insertion and removal.


Hazard / Issue

A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Part Number: 320-35-003 Lot Number: TSL006617 UDI: (01)00812926026893(10)TSL006617
View official government recall