medical Class II Updated 2020-01-22

Flexicare Medical Ltd. recalls BriteBlade Pro Single-Use Fiber Optic Handle and Blade Mille

Recalled Product

BriteBlade Pro Single-Use Fiber Optic Handle and Blade Miller 2, CE 0120, Do not Re-use, Rx Only, REF: 040-342U, Sterile EO, Qty 10, UDI: (01)15055788724053(17)230901(10)181002320


Hazard / Issue

The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Part Number/Lot Number: 040-342U 180802596 I OGH 040-342U 181002320 I OGH 040-342U 181100085 I OGH 040-342U 181200025 I OGH