medical Class II Updated 2022-03-23

GE Healthcare, LLC recalls GE Healthcare CARESCAPE Central Station (CSCS) V2.0x MAI700

Recalled Product

GE Healthcare CARESCAPE Central Station (CSCS) V2.0x MAI700 (Integrated), REF 2082278-001, and MAS700 (Desktop, REF 2082279-001.


Hazard / Issue

If the CARESCAPE Central Station v2.0 is used with an unapproved keyboard, the audio can be muted resulting in loss of audible alarms.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Version 2.0.x (v2.0.0, v2.0.1, v2.0.2, or v2.0.3) with serial numbers beginning with SKN for the MAI700 Integrated, UDI 00840682109666 and serial numbers beginning with SNF for the MAI700 Integrated, UDI 0100840682109604; or serial numbers beginning with SNF for the MAS700 Desktop, UDI 00840682109604.