medical Class III Updated 2024-01-31

Philips North America recalls SmartPath to dStream for 1.5T, Model No. 781260, 782112

Recalled Product

SmartPath to dStream for 1.5T, Model No. 781260, 782112


Hazard / Issue

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model No. 781260 UDI-DI 00884838095076 Serial Numbers 21020 9024 8657 21200 21223 18948 32131; Model No. 782112 UDI-DI 00884838098886 Serial Numbers 30089 20204
View official government recall