medical Class II Updated 2024-02-07

American Contract Systems, Inc. recalls Custom procedural convenience kits and trays, anesthesia, la

Recalled Product

Custom procedural convenience kits and trays, anesthesia, labeled as: a) HEART ANESTHESIA BUNDLE NS, Kit Number LMHA01V; b) OR/Anesthesia Central Line Insertion Kit, Kit Number UIAN43L


Hazard / Issue

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: a) LMHA01V: No UDI/DI, Lot Numbers: 702221, exp. 10/25/223; 654221, exp. 12/12/2023; 954231, exp. 2/15/2024; 903231, exp. 4/7/2024; b) UIAN43L: No UDI/DI, Lot Numbers: 654221, exp. 12/12/2023; 991231, exp. 1/9/2024