medical Class II Updated 2020-01-29

Isopure Corp recalls Isopure Sodium Bicarbonate Mixing and Distribution System (I

Recalled Product

Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3


Hazard / Issue

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Issued by

FDA

Affected States: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, VA
Lot/Code Info: IsoMix Pro UDI: (01)00852456007085; ISOMIX Serial Number IMP-838703 with Pump SN 180210D18
View official government recall