medical Class II Updated 2022-03-30

Dignitana, Inc. recalls DigniCap Delta Scalp Cooling System, Product Number: 900-100

Recalled Product

DigniCap Delta Scalp Cooling System, Product Number: 900-1001, Labeling: The DigniCap Scalp Cooling System User Manual (PN: 0P1DDCD01M-EN_C / IFU 20190325-06-EN), DigniCap Delta Scalp Cooling Wrap (LAB 20190221-05-EN), DigniCap Delta Small Cooling Wrap (LAB 20200616-01-EN, Quick Reference Guide DigniCap Delta (IFU-20190327-02-EN), The DigniCap Scalp Cooling System Training Guide- Global (TR-20200104-05-EN), The DigniCap Scalp Cooling System Training Guide TR-2019-0801-05-US, DigniCap Delta Procedural Expectations- Global (TR-20191028-03-EN), DigniCap Delta Procedural Expectations TR- 20190721-02-US, DigniCap Delta Small Wrap Protocol Global(TR-20200728-02-EN), DigniCap Delta Small Wrap Protocol TR-20200714-01-US, DigniCap Delta Training Tips Global (TR-20200420-04-EN), DigniCap Delta Training Tips TR-20191202-05-US,


Hazard / Issue

Instructions for Use (IFUs) updated to include use of a headband is now recommended for all patients and instructions for the application of the headband. Additionally, IFU updated to include warnings for risk of condensation build up depending on environmental conditions with instructions.

Issued by

FDA

Affected States: LA, NC, NJ, FL, WI, CA, NY, KS, GA, SC, CT, KY
Lot/Code Info: All manufactured units lots/serial numbers: A65860-A73244
View official government recall