medical Class II Updated 2024-02-07

American Contract Systems, Inc. recalls Custom procedural convenience kits and trays, general hospit

Recalled Product

Custom procedural convenience kits and trays, general hospital use, labeled as: a) Baby PICC Kit, kit number UIBP90J; b) Baby PICC Kit, kit number UIBP90K; c) PORT PACK (PS 058330), kit number UIPO82AB; d) Pediatric PICC Kit, kit number UIPP76K;


Hazard / Issue

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: a) UIBP90J, UDI/DI 00191072117038, Lot Numbers: 996231, exp. 1/4/2024; b) UIBP90K, UDI/DI 00191072189356, Lot Numbers: 911231, exp. 7/26/2023; c) UIPO82AB, UDI/DI 00191072172006, Lot Numbers: 727221, exp. 9/30/2023; 680221, exp. 11/16/2023; d) UIPP76K, UDI/DI 00191072142849, Lot Numbers: 637221, exp. 12/29/2023; 976231, exp. 1/24/2024; 898231, exp. 4/12/2024