medical Class II Updated 2021-01-20

Fresenius Medical Care Holdings, Inc. recalls Naturalyte Liquid Acid concentrate bottles, Product Code 08-

Recalled Product

Naturalyte Liquid Acid concentrate bottles, Product Code 08-1251-1 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.


Hazard / Issue

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: 20LXAC063 20NXAC008 Additional Affected lots as of 3/17/21: 20NXAC005 20LXAC098 20NXAC039 20NXAC008 20LXAC063