medical Class I Updated 2025-01-22

Nuwellis Inc recalls nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Numbe

Recalled Product

nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System


Hazard / Issue

The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: 22697, 22698, 22699, 22700, 22701, 22702, 22703, 22704, 22705, 22706, 22707, 22708, 22721, 22722, 22723, 22724, 22733, 22734, 22735, 22736, 22737, 22738, 22739, 22740.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →