Leica Microsystems, Inc. recalls PROVEO 8 Ophthalmic Microscope System optical instrument fo

Recalled Product

PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.


Hazard / Issue

The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.

Issued by

FDA

Distribution: Worldwide
Lot/Code Info: Serial Numbers: 190216001; 060516001; 280916001; 061216001; 091216001; 301015001; 270516001; 180716001; 180716002; 190716001; 141016001; 211016001; 241016001; 261016001; 030317001; 060317001; 011216001; 020317001; 130716002; 100316001; 230216001; 041115001; 021115001; 270416001; 130516001; 160516001; 240516001; 220217001; 100417001; 131115001; 071216001; 100816001; 300816001; 231116001; 121216001; 100816001; 130616001; 081216001; 131216001; 140317001; 310816001; 160816001; 130117001; 060217001; 230317001; 280317001; 050417001; 170117001; 190416001; 070415004; 050516001; 090516001; 170516001; 020816001; 040816001; 111016001; 131116001; 150217001; 211116001; 090217001; 081116001; 020616001; 050117001; 031216001; 020416001; 010317001; 280416001; 271216001; 251115001; 250316001; 231216001; 230916001; 230217001; 260416001; 030616001; 080616001; 080316001; 250116001; 181016001; 220316001; 130716002; 201116001; 200217001; 180416001; 201216001; 170117002; 150116001; 121115001; 110316001; 060616001; 270217001; 090317001; 160317001; 190217001; 140716002; 101016001; 161115001; 190116001; 130116001; 010316001; 020217001; 121016001; 141016001; 161216001; 171216001; 180116001; 260516001; 300516001; 200716001; 260716001; 191116001; 171116001; 251116001; 131216001; 220416001; 281016001; 011116001; 270116001; 060416001; 231115001; 301015003; 301015002; 280116001; 150216001; 180216001; 290416001; 040516001; 190516001; 230516001; 250516001; 280616001; 300616001; 200616001; 230616001; 140716001; 130716001; 010916001; 050916001; 100117001; 260116001; 101116001; 041216001
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →