medical Class II Updated 2021-01-20

Fresenius Medical Care Holdings, Inc. recalls Naturalyte Liquid Acid concentrate drums, Product Code 13-22

Recalled Product

Naturalyte Liquid Acid concentrate drums, Product Code 13-2251-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.


Hazard / Issue

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: 20LXAC025 20LXAC081 20LXAC082 20LXAC083 Additional Affected lots as of 3/17/21: 20LXAC029 20NXAC021 20LXAC026 20LXAC065 20LXAC028 20NXAC029 20LXAC071 20LXAC103 20LXAC027 20LXAC070 20LXAC067 20NXAC019 20LXAC082 20LXAC025 20LXAC081 20LXAC083