medical HIGH RISK Updated 2021-01-27

Penumbra Inc. recalls Penumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Asp

See all recalls from Penumbra Inc. →

Recalled Product

Penumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7KIT The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Product Name: Penumbra JET 7 Reperfusion Catheter and Penumbra Hi-Flow Aspiration Tubing Catalog Number: 5MAXJET7KIT UDI Code: 00815948020962 Lot Number Range: C15441 thru C19176 F88882 thru F100278