medical Class I Updated 2024-02-14

Cardinal Health 200, LLC recalls Cardinal Health Monoject, 35mL Enteral Syringe (purple) - Fo

Recalled Product

Cardinal Health Monoject, 35mL Enteral Syringe (purple) - For Use with ENFit Connection System, REF 435SE


Hazard / Issue

The listed lots of Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) have recognition and compatibility issues with certain syringe enteral feeding pumps. As a result, Cardinal Health recommends that they not be used with enteral syringe feeding pumps.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI/DI 26971564466244 (Case), 16971564466247 (Box),06971564466240 (Each), Lot Numbers: 230501, 230601 and 230602