Shimadzu Medical Systems recalls MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
Recalled Product
MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
Hazard / Issue
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
Issued by
FDA
Affected States: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WV
Lot/Code Info: Model: FLUOROspeed X1 Generator Number: D150-40 Serial Numbers: MQ493E19C001 MQ493E3B8002 MQ493E3B9001 MQ493E1B4001 MQ493E1A1003 MQ493E1A3001 MQ493E1B5001 MQ493E3B8001 MQ493E1A5001 MQ493E1A1005 MQ493E1A2001 MQ493E1A4001 MQ493E1B6002 MQ493E1A8001 MQ493E1B4002 MQ493E1A4003 MQ493E1A4002 MQ493E1B6001 MQ493E1A2002 MQ493E1AC001 MQ493E1A1002 MQ493E19C002 MQ493E1A3003 MQ493E1A1004 MQ493E1A3002