medical Class II Updated 2021-01-27

The Magstim Company Limited recalls Neurosign 100, Product Number 9883-01. Intraoperative Nerve

Recalled Product

Neurosign 100, Product Number 9883-01. Intraoperative Nerve Monitor - Product Usage: indicated for use in stimulating and monitoring cranial motor nerves, especially the facial nerve (VII), during ENT surgical procedures.


Hazard / Issue

A small number of Neurosign 100 Intraoperative Nerve Monitors may have been dispatched without being configured appropriately for the supply mains within the target Country. The specific deficiency includes: 1. Onboard voltage selector set to 240V (European Standard), versus 120V requirement for the United States and Canada. 2. Fuses installed in the Power Entry Module (PEM) 2xT315mAL (240V Standard), versus the 2xT630mAL requirement for a 120V mains supply.

Issued by

FDA

Affected States: MN, NC, NY
Lot/Code Info: Serial Numbers: 2507, 2511, 2514, 2515, 2522, 2527, 2533, 2534, 2537