Shimadzu Medical Systems recalls MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
Recalled Product
MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
Hazard / Issue
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
Issued by
FDA
Affected States: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WV
Lot/Code Info: Model: SonialVision safire: Generator Number: D150-40 Serial Numbers: 3M5249D41006 3M5249D2A004 3M5249D34001 3M5249D37006 3M5249D32012 3M5249D35002 3M5249D1C013 3M5249D31009 3M5249D29002 3M5249D33003 3M5249D2C004 3M5249D22013