Philips Medical Systems Nederlands recalls Multiva 1.5T 16 R5 (ZBO), Model Number 781073. Nuclear Mag
Recalled Product
Multiva 1.5T 16 R5 (ZBO), Model Number 781073. Nuclear Magnetic Resonance Imaging System
Hazard / Issue
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Issued by
FDA
Distribution: Nationwide (United States)
Lot/Code Info: All systems