medical Class II Updated 2021-01-27

Abbott Point Of Care Inc. recalls Abbott i-STAT CG8+ cartridges - IVD Intended for quantitativ

Recalled Product

Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25


Hazard / Issue

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Stored at room temperature (18 to 30¿C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)