medical Class II Updated 2021-01-27

Abbott Point Of Care Inc. recalls Abbott i-STAT EG7+ cartridge-IVD. Intended for the quantita

Recalled Product

Abbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood. List Number: 03P76-25


Hazard / Issue

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Stored at room temperature (18 to 30¿C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →