medical Class II Updated 2020-02-05

Haemonetics Corporation recalls High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device -

Recalled Product

High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.


Hazard / Issue

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog number/FG Item Number: 0263A-00 1. No lots for cracking issue identified (0 Distributed). 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.