medical Class II Updated 2024-02-07

Howmedica Osteonics Corp. recalls SCORPIO U-DOME PATELLA -Intended for knee replacement

Recalled Product

SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3110


Hazard / Issue

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GTIN: 07613327033861 Lot Numbers: XEA7, 8P83, 690Y, MRM3, 67D4, 4902, YNP7, 3AYN, AYHM, W3MD, 76KN, K4P9, 792H, 43T8, MW40
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →