medical Class I Updated 2025-01-22

Olympus Corporation of the Americas recalls Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Mod

Recalled Product

Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.


Hazard / Issue

Potential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number: MAJ-891. UDI-DI: 04953170063114. All lot numbers.