medical Class II Updated 2025-01-22

CareFusion 303, Inc. recalls BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxi

Recalled Product

BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352


Hazard / Issue

Due to an automated dispensing cabinet software issue, when a medication single order contains more than one component and at least one is unavailable for a multi-item or combination order, the order may display as grayed out; alternate location tooltip will display "Server connection failed," not listing other area devices where unavailable item(s) may be found, which may lead to delayed therapy.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Software v1.7.3. REF/UDI-DI: 323/10885403512667, 352/10885403512674
View official government recall