medical Class II Updated 2021-02-03

Respironics California, LLC recalls Philips Respironics V60 Ventilator Part Number R1053618 - Pr

Recalled Product

Philips Respironics V60 Ventilator Part Number R1053618 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.


Hazard / Issue

Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model: V60 Part Number: R1053618 V60 Ventilator, Japan Opt: CFLEX,AVAPS Serial Numbers: 100039155 100054624 100054738
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →