medical Class II Updated 2020-02-05

Beckman Coulter Inc. recalls iChemVELOCITY, Urine Chemistry System, Catalog numbers: 700

Recalled Product

iChemVELOCITY, Urine Chemistry System, Catalog numbers: 700-7176-001, 700-7177-001, B75502 - Product Usage: used to automate the urine chemistry analysis profile using iChem¿ VELOCITY Urine Chemistry Strips.


Hazard / Issue

Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics).

Issued by

FDA

Distribution: Worldwide distributions.
Lot/Code Info: UDI: 10837461009375, 10837461001751; All Part Numbers