medical MODERATE Updated 2017-01-04

Siemens Medical Solutions USA, Inc recalls SOMATOM Scope Power, Computed tomography x-ray system

See all recalls from Siemens Medical Solutions USA, Inc →

Recalled Product

SOMATOM Scope Power, Computed tomography x-ray system

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.

Issued by

FDA

Affected States: IL, GA, SC, SD, OK, NY, MI, IN, MD, AR
Lot/Code Info: Model # 10967888 Serial Number:59376 91766 91767 115456 115454 59380 77560 98840 77557 91758 59377 59378 115451