medical Class I Updated 2024-02-14

Philips North America recalls BrightView X, gamma camera for SPECT; Model Nos.: 6-di

Recalled Product

BrightView X, gamma camera for SPECT; Model Nos.: 6-digit format 882478; 12-digit format 453560824741 453560829261;


Hazard / Issue

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All serial numbers are affected. No UDI-DI available for this product.
View official government recall