medical Class II Updated 2022-04-20

Graphic Controls Acquisition Corporation recalls CLARAVUE Pre-wired Electrode - IP SET 5V RT ADULT Model:

Recalled Product

CLARAVUE Pre-wired Electrode - IP SET 5V RT ADULT Model: 32028781 50604-US


Hazard / Issue

Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots UDI: 03700506307302