medical Class II Updated 2019-03-06

Arrow International Inc recalls AF550x5ML/H 2ML 60LO 6" DUAL CATHS, 4.5", Product MT5060XLYK

Recalled Product

AF550x5ML/H 2ML 60LO 6" DUAL CATHS, 4.5", Product MT5060XLYK5 CP, Arrow AutoFuser Disposable Pain Control Pump


Hazard / Issue

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 74F1700794 74J1700676 74J1700190
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →