medical Class I Updated 2020-02-19

Medtronic Inc. recalls Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s)

Recalled Product

Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and MMT-1752 *Not for US Distribution


Hazard / Issue

There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.

Issued by

FDA

Affected States: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Lot/Code Info: All Serial Numbers ( excluding devices with black retainer rings) ***NOT DISTRIBUTED WITHIN THE US***