medical Class II Updated 2020-02-12

Cook Inc. recalls Bentson PTFE Wire Guide, Guidewire, Global Product Number G1

Recalled Product

Bentson PTFE Wire Guide, Guidewire, Global Product Number G14589, G14590


Hazard / Issue

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 7832513 7832514 8062406 8062407 8065988 8065989 8065992 8404354 8404355 8547811 8069648