medical Class II Updated 2022-04-27

Siemens Medical Solutions USA, Inc recalls Artis zee floor MN, Model Number 10094142

Recalled Product

Artis zee floor MN, Model Number 10094142


Hazard / Issue

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI: 4056869010090; Serial Numbers: 140421 140115 140101 140205 140213 140204 140209 140219 140224 140218 140226 140221 140220 140206 140105 140423 140406 140422 140102 140214 140109 140202 140424 140114