medical Class II Updated 2022-04-27

Siemens Medical Solutions USA, Inc recalls Artis zee biplane MN, Model Number 10094143

Recalled Product

Artis zee biplane MN, Model Number 10094143


Hazard / Issue

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402
View official government recall