medical Class I Updated 2019-04-10

O-Two Medical Technologies, Inc. recalls e500 Automatic Transport Ventilator, Model Number 01EVE500

Recalled Product

e500 Automatic Transport Ventilator, Model Number 01EVE500


Hazard / Issue

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Issued by

FDA

Affected States: AZ, CA, CO, FL, OH, PA, TX, VA
Lot/Code Info: EV50001 to EV50148
View official government recall