ZOLL Medical Corporation recalls OneStep CPR Complete, Adult Multi Function Electrodes, Part
Recalled Product
OneStep CPR Complete, Adult Multi Function Electrodes, Part Numbers: 8900 0214 01 (case of 8 single electrodes, UDI: 10847946016279) and 8900 0224 01 (single electrode, UDI: 00847946016272) Product Usage: Intended Use: Defibrillation, Cardioversion, Noninvasive Pacing, EGG Monitoring, CPR Sensor. For use with ZOLL¿ Defibrillators: R Series, M Series. By Trained Personnel only, Including: Physicians, Nurses, Paramedics, Emergency Medical Technicians, Cardiovascular Laboratory Technicians.
Hazard / Issue
A portion of one lot of ZOLL OneStep Complete Electrodes for adults will cause the defibrillator to deliver pediatric energy levels instead of adult energy levels when used with the ZOLL R Series Defibrillator. Defective electrodes will display PEDIATRIC PADS IN USE on the R Series defibrillator. Using the affected product with the M Series Defibrillator poses no risk.