medical Class I Updated 2019-04-10

O-Two Medical Technologies, Inc. recalls e700 Automatic Transport Ventilator, Model Number 01EVE700

Recalled Product

e700 Automatic Transport Ventilator, Model Number 01EVE700


Hazard / Issue

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Issued by

FDA

Affected States: AZ, CA, CO, FL, OH, PA, TX, VA
Lot/Code Info: EV70001 to EV70658