medical Class II Updated 2020-02-12

Cardinal Health 200, LLC recalls Genius 2 Tympanic Thermometer OEM Tympanic Item Code:3030

Recalled Product

Genius 2 Tympanic Thermometer OEM Tympanic Item Code:303062


Hazard / Issue

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All product manufactured after October 1, 2016; Serial Numbers greater than or equal 16597907