Randox Laboratories Ltd. recalls RX Series Copper (Cu) Assay Ref. Number CU2340
Recalled Product
RX Series Copper (Cu) Assay Ref. Number CU2340
Hazard / Issue
Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results.