medical Class II Updated 2024-02-14

Randox Laboratories Ltd. recalls RX Series Copper (Cu) Assay Ref. Number CU2340

Recalled Product

RX Series Copper (Cu) Assay Ref. Number CU2340


Hazard / Issue

Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalogue Number/GTIN: CU2340 05055273201949; Lot/batch: 592278, 610661, 610669, 629021, 647201, 657944;