Recalled Product

PrisMax System, Product Codes: a) 955558, b) 955627, c) 955701


Hazard / Issue

This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: a) 955558, UDI 07332414124359, b) 955627, UDI 07332414124700, c) 955701, UDI 07332414125844, All Serial Numbers