medical Class II Updated 2022-05-04

Intuitive Surgical, Inc. recalls da Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci

Recalled Product

da Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi Sureform 45 Curved-Tip REF 480545 da Vinici X da Vinci Xi Sureform 60 REF 480460


Hazard / Issue

As a result of complaint data review, it identified tissue pushout events occurring during firing of the SureForm stapler when tissue is pushed distally within the jaws, rather than being held in place and transected. Overall rate of harm associated with bleeding of critical severity exceeds the predicted likelihood of harm as documented within Intuitives existing risk documentation.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Product Description/Model Numbers/UDI Code: Sureform 45 / 480445 / 00886874117583 Sureform 45 Curved-Tip / 480545 / 00886874117590 Sureform 60 / 480460 / 00886874115640 All Lots
View official government recall

Looking for safe alternatives?