medical Class II Updated 2024-02-14

Medtronic MiniMed, Inc. recalls Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7

Recalled Product

Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7040C2, MMT-7040C3, MMT-7040D1, MMT-7040D2, MMT-7040MA, MMT-7040QC1, MMT-7040QC2, MMT-7040QC3


Hazard / Issue

Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 ¿m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function due to the sensor layer being too thin. The potential impact resulting from this issue ranges from user inconvenience to over/under delivery of insulin, possibly resulting in hyperglycemia or hypoglycemia, if the sensor glucose values are affected. The issue was identified by the manufacturing team during the execution of Filmetrics Measurement for a Guardian Sensor lot.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: REF/UDI-DI/Lot: MMT-7040A/20763000413389/HG72B2C, HG72CQQ, HG73VB7, HG73VC8, HG73VFT, HG73WFR, HG7441E, HG745F9, HG7486X, HG74AJ1, HG74SUR, HG74T9Q, HG74U36, HG78LMS, HG79Y9U; MMT-7040C1/00763000555276, 00763000413392, 20763000413396, 20763000555270, 20763000712475, 00763000712471/F1323, F1923, G0323, G1723, G2023, G2623, G2923, HG6RV7S, HG6TGNU, HG6TJ2Y, HG6TW97, HG6TWVU, HG6UNF2, HG6V8NN, HG6VB8S, HG6VNX3, HG6WNE1, HG6XFPR, HG6XH54, HG6XQCG, HG6XZJX, HG6Y0W1, HG6Y3QN, HG6YEK6, HG6YFAF, HG6YFHL, HG6YJAR, HG6YJBD, HG6YNGD, HG6Z7N5, HG6ZARN, HG7038Y, HG70CW5, HG70DAM, HG70JSU, HG70LRC, HG70QWA, HG71VMZ, HG72DB5, HG72G9C, HG72NV9, HG72R1P, HG72RBD, HG72SE3, HG72TPT, HG72TRV, HG72V3J, HG72ZFP, HG7340A, HG73EJX, HG73RNR, HG745SV, HG747E3, HG7480Q, HG749N6, HG74F93, HG75TSX, HG75XA1, HG76HFH, HG76RYY; MMT-7040C2/20763000555287/HG6RC39, HG6RXB5, HG6T4T7, HG6T8RV, HG6TJ2Y, HG6WKZU, HG6WLKS, HG6XFPR, HG6XH54, HG6XLSD, HG6XNH3, HG6Y5C6, HG6Z1X5, HG713AK, HG73EGX, HG73EJX, HG73FJP, HG73Q24, HG73QPM, HG73RNR, HG7619X, HG764J9, HG76CWC, HG77FMK; MMT-7040C3/20763000555294/HG6XNH3, HG73RNR; MMT-7040D1/00763000413484/HG6THY1, HG6V7VV; MMT-7040D2/00763000413491/HG6TL44; MMT-7040MA/20763000519692/HG74460; MMT-7040QC1/00763000555313, 20763000555317/HG6WT9A, HG6WUGE, HG73PH3; MMT-7040QC2/20763000555324, 763000555320/HG6WT9A; MMT-7040QC3/20763000555331/HG6WT9A
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →