Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista 500, Analyzer, chemistry Model no. 10284473
Recalled Product
Dimension Vista 500, Analyzer, chemistry Model no. 10284473
Hazard / Issue
When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination.
Issued by
FDA
Distribution: Nationwide (United States)
Lot/Code Info: Millipore lot number: F7MA50339 F7MA56608 F7NA62488 F8AA13966 F8AA93567 F8BA20457 F8BA35774 F8DA59251 F8HA64477 F8HA79819 F8HA91305 F8JA97876 F8MA06374