medical Class II Updated 2022-05-04

Instrumentation Laboratory recalls ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part N

Recalled Product

ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. ACL TOP 500 CTS 00000280040 5. ACL TOP 500 CTS 00000280040R ( Refurbished) 6/ 6. ACL TOP 550 CTS 00000280045 7. ACL TOP 970 CL 00000280097


Hazard / Issue

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All Serial Numbers. Trade Name Part Number UDI ACL TOP 300 CTS 00000280060 08426950556916; "ACL TOP 300 CTS (Refurbished)" 00000280060R 08426950928027; ACL TOP 350 CTS 00000280065 08426950784081; ACL TOP 500 CTS 00000280040 08426950453499; "ACL TOP 500 CTS (Refurbished)" 00000280040R 08426950928003; ACL TOP 550 CTS 00000280045 08426950729242; ACL TOP 970 CL 00000280097 08430793045476.