medical Class II Updated 2019-03-27

Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista 1500, Analyzer, chemistry Model no. 104448

Recalled Product

Dimension Vista 1500, Analyzer, chemistry Model no. 10444801


Hazard / Issue

When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Millipore lot number: F7MA50339 F7MA56608 F7NA62488 F8AA13966 F8AA93567 F8BA20457 F8BA35774 F8DA59251 F8HA64477 F8HA79819 F8HA91305 F8JA97876 F8MA06374
View official government recall