medical Class II Updated 2017-02-01

On-X Life Technologies, Inc. recalls Chord-X Chordal Sizer is designed to assist in determining t

Recalled Product

Chord-X Chordal Sizer is designed to assist in determining the proper length of chordal replacement prostheses to repair the mitral valve and help the surgeon select a preconfigured prosthesis, or configure the correct size using straight suture material, during mitral valve repair.


Hazard / Issue

Four issues were uncovered by the recalling firm's CAPA: 1) Burrs were found to be present on Left Housing, Right Housing and Barrel components, 2) Loose plastic particulate was found within sealed primary packages, 3) Loose hair was discovered within four sealed primary packages and 4) Assembly error, wherein Left and Right Housing components were acceptably press-fit together, was discovered within one sealed primary package. One of the Left Housing press-fit pins was not pressed into the associated hole of the Right Housing, resulting in a small gap in the housing assembly.

Issued by

FDA

Affected States: CA, WA, NE, DE, OH, ME, TX
Lot/Code Info: 72-00-47420
View official government recall