medical Class II Updated 2019-04-03

Keystone Dental Inc recalls Paltop Premium Surgical Kit-containing the 1.25mm Hex Driver

Recalled Product

Paltop Premium Surgical Kit-containing the 1.25mm Hex Driver For surgical and restorative applications prosthetic devices, such as artificial teeth Part Number: 60-70112


Hazard / Issue

Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature

Issued by

FDA

Affected States: FL, MA, SC, VA
Lot/Code Info: Lot Numbers: WO-010252, WO-010136
View official government recall