medical Class II Updated 2022-05-18

Karl Storz Endoscopy recalls 11272VH-TL 11272VHK-TL HD-VIEW Flexible HD Cysto-Urethroscop

Recalled Product

11272VH-TL 11272VHK-TL HD-VIEW Flexible HD Cysto-Urethroscope Z23875US-BC (10-2021) 11272VHU-TL 11272VHUK-TL HD-VIEW Flexible HD Cysto-Urethroscope Z23875US-BC (10-2021)


Hazard / Issue

Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.

Issued by

FDA

Affected States: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Lot/Code Info: 11272VH-TL 11272VHK-TL HD-VIEW Flexible HD Cysto-Urethroscope Z23875US-BC (10-2021) UDI Code: 4048551449677; 11272VHU-TL 11272VHUK-TL HD-VIEW Flexible HD Cysto-Urethroscope Z23875US-BC (10-2021) UDI Code: 4048551449684; All Serial Numbers manufactured/distributed since January 2018.
View official government recall