medical Class I Updated 2025-02-26

Spectranetics Corporation recalls Intact Vascular Tack Endovascular System, to treat vascular

Recalled Product

Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 156080062


Hazard / Issue

Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: REF/UDI-DI/Batch: 154150041/00863328000189/332905 - 349049, 154150042/328641, REF: 154090042/322101, 156135061/00863328000103/332824 - 336121, 206135061/00850003494043/331281 - 336152, 206135062/328643 - 345567, 156135062/271330 - 283066, 206080062/321083 - 349035, 156080062/302251 - 349034
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →